COMPARATIVE STUDY OF THE FATIGUE STRENGTH OF DENTAL IMPLANT SYSTEMS WITH PREANGLED AND STRAIGHT CONNECTING PARTS

J.M. Ayllón  J. Mitjavila  C. Navarro  J. Domínguez     

Anales de la Mecánica de la Fractura, nº 33 . 2016 . Pág. 51 -56
Ver (.pdf): Anales33-007

Resumen: Dental implants need to pass a certification where their fatigue strength is determined following the international standard ISO 14801. This standard distinguishes between straight and pre-angled connecting parts. In some countries, it is obligatory to obtain the fatigue curve for the pre-angled implant although it may have already been obtained for the straight one. This paper shows experimentally and analytically that the implant with pre-angled connecting part is more resistant that the straight implant. Furthermore, it is shown that only one type of implant, either straight or pre-angled, needs to be tested and from this result the fatigue curve of the other one can be inferred. For this purpose, various sets of implants, straight and pre-angled for two different designs, have been tested, showing the fatigue lives against an equivalent stress. Finally, the effect over fatigue life of different geometric parameters has been analysed.

LocalizaciónSan Sebastián

Departamento de Ingeniería Mecánica y de Fabricación, Escuela Superior de Ingenieros. Camino de los Descubrimientos s/n, 41092, Sevilla, España.
Departamento de Ingeniería Mecánica y de Fabricación, Escuela Superior de Ingenieros. Camino de los Descubrimientos s/n, 41092, Sevilla, España.
Departamento de Ingeniería Mecánica y de Fabricación, Escuela Superior de Ingenieros. Camino de los Descubrimientos s/n, 41092, Sevilla, España.
Departamento de Ingeniería Mecánica y de Fabricación, Escuela Superior de Ingenieros. Camino de los Descubrimientos s/n, 41092, Sevilla, España.